Respiratory Gating System AZ-733VI

 

Respiratory Gating System AZ-733VI

Overview

What is respiratory gating?

The problem with performing image diagnosis and radiotherapy of tumors arising in organs affected by respiratory motion such as the lungs and liver is that the position of the tumor moves due to the patient’s breathing during imaging or radiotherapy.

These respiratory movements may cause the resolution of the captured image artifacts occur. And the result it may affect the diagnosis. In addition, during radiation therapy, there is concern that the efficacy will be reduced as the tumor may not receive a sufficient dose of radiation, and that surrounding healthy tissue will be irradiated instead, resulting in the side effect of undesirable radiation damage. It is known that steady breathing can be approximated by periodic repetitive movements. With this knowledge, it is possible to take advantage of specific respiration states (e.g. maximum inhalation, complete exhalation, etc.). If imaging or irradiation is performed only in the stationary periods between respiration states, it is possible to achieve high-precision imaging and radiation therapy without the negative effects of respiratory motion. In this way, the respiratory gating system is a medical device that performs respiratory gating imaging and respiratory gating irradiation in accordance with respiratory motion. The respiration gating system uses a sensor to detect a signal representing the respiration state, creates a respiration signal (or a respiratory signal that is synchronized with the respiration signal), and supplies it to a diagnostic imaging or radiotherapy device. These diagnostic imaging and radiotherapy devices perform respiratory synchronized imaging and irradiation by performing imaging and irradiation in concordance with the respiratory signal or respiratory signal. Currently, respiratory gating systems are commonly used alongside CT devices are to record 4D-CT image throughout the entire respiratory cycle of a patient. This image is then used to create a treatment plan for the patient’s radiation therapy.

What is Anzai Medical’s Respiratory Gating System?

It has been a quarter of a century since the world’s first respiratory gating system was developed by us in collaboration with Tsukuba University Hospital in 1987. In the years since, Anzai Medical has perfected our Respiratory Gating System AZ-733VI.

In response to demand for faster respiratory gating systems in the radiotherapy field, we have achieved a respiratory signal output delay of 50 ms or less between detection of respiratory signals and respiratory signal output. As a breathing sensor, in addition to the load cell (contact) that has been standard component; we have prepared a laser sensor option (non-contact) to flexibly respond to various requests in the medical field. Also included is the functionality to connect to and acquire the signals of 3 respiration sensors at the same time. Functionally enhanced is the ability to check and adjust all parts of the respiration sensor on the patient from a nearby port, greatly reducing the time required for a previously laborious process. In addition to various safety functions inherited and enhanced from the previous model AZ-733V, the sophisticated user interface on the AZ-733VI’s personal computer screen enables intuitive operation and provides the ability to pre-check settings to prevent erroneous operation. In addition, a gate disable switch is equipped as standard to respond immediately to sudden changes in the patient’s breathing condition. In order to flexibly support connection with external devices such as diagnostic imaging and radiotherapy equipment, up to three interface boards are mounted on the AZ-733VI (two on the sensor port and one on the relay box). The gate signal is output to only one device which is selected on the personal computer screen. A wide range of external devices are compatible in combination with various respiratory signal output modes further enhanced from the previous model AZ-733V.
Dr. Takashi Ogino Director, International Proton Therapy Center Connection Device: Proton therapy device (Mitsubishi Electric) Purpose of use: Respiratory gated irradiation of lung and liver tumors

Installation

More than 2000 units Anzai Medical’s respiratory gating system AZ-733VI have been shipped to all over the world. And it can be used in combination with the radiotherapy equipment and diagnostic imaging equipment of the following companies.

CT

PET-CT

Linear Accelerator

Particle Therapy

Features

High Performance

 

Scalability

 

Convenience

 

Efficiency

 

Safety Function

 

High Performance

Respiratory Gating System AZ-733VI has devolved the basic performance required for a respiratory synchronization device.

Respiratory sensor to be used according to modality

Two types of high-accuracy respiratory sensors can be used according to the modality to be combined.

Loadcell Sensor (patient contact style)
ロードセルセンサ(患者接触方式)

It is a sensor that detects pressure with a strain gauge. By attaching the sensor to the patient with a belt, the movement of the body surface (change in abdominal pressure) due to respiratory movement is displayed as a respiratory waveform. “DEEP” and “STANDARD” are used according to the size of the abdominal movement. “STANDRAD” is used for standard movements, and “DEEP” is used for large movements. As a modality to be combined, it is ideal for PET-CT devices that allow patients to enter deep inside the gantry. For the setting of this sensor, see “How to install the load cell sensor” and “How to calibrate load cell sensor”.

Laser Sensor (patient non-contact style)
レーザセンサ(患者非接触方式)

It is a sensor that measures the reflection of laser light (infrared). A laser beam is applied to the patient’s skin and the movement of the body surface (displacement amount to the sensor) due to respiratory movement is displayed as a respiratory waveform. We have three types of 250mm, 120mm, and 85mm depending on the focal length of the laser. This sensor, which does not come into contact with the patient, is best suited as a modality to be combined with radiation therapy equipment.For the setting of this sensor, see “How to set the laser sensor” and “How to calibrate load cell sensor”.

Reduced delay time
遅延時間*の短縮グラフ

High-speed gate mode: within 25 msec Normal gate mode: Within 50 msec

Respiratory gating system is used in combination with radiotherapy equipment and image diagnostic equipment. The biggest subject to reduce the performance for modalities being combined was to reduce latency. Compared to conventional equipment, the speed is reduced to 1/4 in high-speed gate mode and 1/2 in normal gate mode. This makes it possible to realize more accurate radiation treatment and image diagnosis than ever before. The delay time is the time from when the sensor detects the movement of the body surface due to respiratory movement until the gate signal is output.

Scalability

Respiratory gating system AZ-733VI has new functions and connections that have never been seen before.

Excellent cost performance
Conventionally, one external device was connected to one device, but AZ-733VI can connect up to three types of external devices to one device. This allows the same respiratory synchronization system to be used when migrating from simulated CT to treatment, enabling low budget implementation. 1台に最大3種の外部装置接続 ソフトウェア上でゲート信号出力先の装置を選択画面 Select the device to output the gate signal on the software. *Note that the gate signals cannot be output simultaneously.
Using multiple sensors
Respiratory gating system AZ-733VI can use up to 3 respiratory sensors simultaneously. Since the movement to the body surface due to respiratory movements differs depending on the patient, it is possible to capture respiratory information from multiple locations and perform respiratory gating at the most stable location.
複数センサの使用
Respiratory waveform 3 types displayed 呼吸波形図
Output of ECG/breath gating signals
心電呼吸波形同期信号図

Respiratory gating system AZ-733VI has a function to output an electrocardiographic/respiratory gating signal by connecting an electrocardiograph. When the arbitrarily set respiratory gating signal and electrocardiographic gating signal overlap, the electrocardiogram/respiratory gating signal is output to an external device.

Convenience

By improving operability and fulfilling file management, we aimed to create an application that will satisfy users all over the world.

Easy respiratory waveform adjustment
センサポート

By operating the rotary encoder mounted on the sensor port, the respiratory waveform can be adjusted near the patient. It is no longer necessary to return to the operation room and adjust the waveform from the software after installing the sensor as in the past. In addition, the waveform is displayed on the LCD screen of the sensor port, which improves reproducibility when mounting the sensor.

Extensive patient data management

In addition to saving and replaying the saved patient data, the respiratory information is displayed in time series to facilitate understanding of the patient’s respiratory characteristics, and a histogram is displayed by aggregating by respiratory level or respiratory rate.

Table display of patient data 患者データのテーブル表示

Histogram display of patient data 患者データのヒストグラム表示

Efficiency

The efficiency of respiratory gating (short-time treatment/photographing) reduces the burden on patients.

Addition of new gate output mode
A new “Level to Level” mode, which is mainly used for radiation therapy, has been added. The gate signal is output to an external device by deciding the phase that is less affected (moved) by the respiratory phase. Since the gate signal is output within the set level regardless of expiration or inspiration, breath-hold irradiation is also possible.
『Level to Level』モード The above outputs the gate signal at Level 20-60, and two gate signals are output during one respiratory cycle.
Addition of breath monitor
呼吸モニタ

Respiratory gating is a treatment/imaging that depends on the patient’s respiratory condition. We have prepared a respiratory monitor so that the patient can see the respiratory waveform and consciously breathe. It is hoped that the patient’s breathing condition will be stable and the treatment/diagnosis time will be shortened.

Safty Function

It is available to use Respiratory Gating System AZ-733VI safely with its various safety functions.

Stop gate signal in case of emergency
In an emergency, you can use the gate cutoff switch to forcibly stop the gate signal output.
Abnormal breath detection
When the following breathing abnormalities are detected, an error message is displayed on the software and the gate output to the external device is stopped.

Specification

Sensor port

センサーポート
Dimension : 260 x 230 x 126 mm
Weight : about 5.0kg
Power supply : AC100-240V (50/60Hz), 200VA or less
Monitor : Respiratory waveform information display, connected sensor display
Others : Automatic adjustment of respiratory waveform of connected respiratory sensor

Relay box

中継ボックス
Dimension : 172 x 123 x 86 mm
Weight : about 2.0kg
Output signal : Gate signal, (beam signal, waveform information, interlock)

Load Cell (2 types) STANDARD, DEEP

ロードセル(2種) STANDARD、DEEP
Dimensions Sensor part : 30φ x 9.5, load cell amplifier: 85 x 27 x 30 mm
Weight : about 95g
Measuring range : 0-500g (STANDARD), 0-1000g (DEEP)
Cable length : 2.9m

Gate cutoff switch

ゲート遮断スイッチ
Dimension :70×35×50mm
Weight :about 0.3kg

Personal computer

パーソナルコンピュータ
Power supply :AC100-240V (50/60Hz), 200VA or less
OS :Windows 10
Dimension :384×34×258
Weight :about 3kg

Laser sensor (3 options) 250mm, 120mm, 85mm

レーザセンサ(オプション 3種) 250mm、120mm、85mm
Dimensions Sensor part : 67 x 22 x 57 mm, Laser sensor amplifier: 65 x 160 x 60 mm
Weight :about 360g
Measuring range (250mm) :250mm±150mm
(120mm) :120mm±60mm
(85mm) :85mm±20mm

ABLE (option)

ABLE(オプション)
Power supply voltage tablet: battery-powered transceiver: DC+12V (powered by AZ-733VI
Transmission/reception frequency band Tablet: 2.4GHz band Transceiver: 2402 to 2480MHz
Dimensions Tablet: 83 x 159 x 10 mm Transceiver: 85 x 35 x 25 mm
Weight Tablet: Approx. 200 g Transceiver: Approx. 80 g

Sensor Port Stand (option)

Sensor Port Stand (option)
Dimensions : Φ520×800mm
Weight :about 8.7kg

Licence

  • JAPAN
    Medical device approval number : 22600BZX00288000
    Class classification : Class III
    General name : Synchronizer for radiotherapy equipment
    Specified maintenance management medical device : Applicable
    nstallation management medical device I: Applicable
    JMDN code

Agency

  • USA Company Name: XRT Medical LLC Address: 2900 Fiore Way, #112, Delray Beach, FL 33445 USA, HP:https://xrtmedical.com/
  • EU Company Name:ELSE Solutions s.r.l. Address:  Via Carlo Goldoni, 18 20090 Trezzano Sul Naviglio (MI) – Italy HP:https://www.elsesolutions.com/
  • Switzerland Company Name:  Raditec Medical AG Address:  Schlossberg 5a CH-5454 Bellikon Switzerland HP:https://www.raditec.ch/
  • China Company Name: Beijing Boiatech Medical Technology Co.,Ltd Address: Room 2-2146, Building 2, Century Tea Trade Center, No.1 Chama North Street, Xicheng District, Beijing, China (100055) HP:http://en.boiatech.com.cn/
  • Taiwan Company Name: Chili Enterprise Co., Ltd. Address:  18f-12, No. 495, Guangfu S Rd., Xinyi Dist., Taipei City 110, Taiwan (R.O.C.) HP : https://chillimedtech.com/
  • Thailand Company Name: CMC BIOTECH CO., LTD. Address:  364 Soi Ladphrao 94(Panjamitr), Ladphrao Road, Plab-Pla,Wangthonglang, Bangkok 10310, Thailand HP:http://www.cmcbiotech.co.th/th/
  • Malaysia Company Name: H&A MEDICAL Address: CT-05-11 Corporate Tower Subang Square, Jalan SS15/4G, Subang Jaya, 47500 Selangor, HP:https://hnamedical.com/
  • Saudi Arabia Company Name: ASCO Address: Riyadh Po. Box 220646 – 11311 HP:https://www.ascomed.com
  • Latvia Company Name: A.MEDICAL LLC Address: Varkaļu str. 13a, Rīga, LV-1067 HP:http://www.amedical.eu/
  • Colombia Company Name: TOP MEDICAL SYSTEMS S.A Address: Cra. 18A No. 103 – 47 / Bogota – Colombia HP:http://topmedical.com.co/
TOP
HOME
Inquiry