安西メディカル株式会社

Frequently Asked Questions

Q1
Is the AZ-M1PP a medical device?
A
Yes, it is a certified medical device.
Q2
What isAZ-M1PP used for?
A
It is a treatment QC phantom designed to simulate the human body, consisting of bone, lung vessels, and a simulated tumor, used as a tracking chest motion phantom.
Q3
How does the phantom operate?
A
The simulated tumor and lung vessels move independently. The movement range is ±25 mm, and it can operate using selectable waveforms such as sine, cos4, cos6, triangle wave, or custom waveforms (via AZ-733VI). The positioning accuracy is ±0.2 mm.
Q4
How does the body surface unit operate?
A
The body surface unit moves within a range of ±10 mm and can be operated using selectable waveforms such as sine, cos4, cos6, triangle wave, or custom waveforms via AZ-733VI. It has a positioning accuracy of ±0.2 mm.
Q5
Does the device operate using respiratory waveforms acquired with AZ-733VI?
A
Yes, it imports waveforms in arbitrary waveform input mode and operates based on the acquired respiratory data.
Q6
Is manual operation possible?
A
Yes, the device is equipped with Wi-Fi, allowing operation from Wi-Fi-enabled computers, tablets, and smartphones. After connecting to Wi-Fi, you can download dedicated software to control the device manually.