Frequently Asked Questions

Q
Which is more accurate, load cell or laser sensor?
A
There is no difference in sensitivity, so use it according to your application.The load cell that is directly attached to the patient's body is suitable for PET-CT with a deep gantry.The non-contact laser sensor, which is not directly attached to the patient, is suitable for use in radiation therapy and fixtures.
Q
What is the difference between load cell standard and deep?
A
The standard has a higher sensitivity and the measuring range has a wider measuring range.The standard is used for patients with relatively small body movements due to breathing.Deep is used for patients with large body movements due to breathing.
Q
No test waveform is output.
A
It is necessary to enable "TEST SIGNAL" from the software setting.Also, all components such as sensors must be connected.If "Test Signal" is displayed in the center of the screen of the application but the test waveform is not displayed, adjust the Magnification/Position.
Q
How can I connect multiple external devices to one respiratory sync system AZ-733VI?
A
You will need to add a dedicated cable and dedicated interface for the device to be connected.Connection cables may be added depending on the layout of the room used.
Q
What type of the cables are used?
A
Please refer to the connection layout below.
QA image
Q
How often is the maintenance required?
A
We recommend annual low-voltage power supply inspection, detector performance test, comprehensive operation test (charged).
Q
How long is the warranty period?
A
The warranty period is one and a half year after delivery, but there is a 1 to 5 year extended warranty service (charged) if necessary.