安西メディカル株式会社
Dynamic Chest Tracking Phantom

Overview

Chest Phantom for motion tracking AZ-M2 is composed of bones, pulmonary blood vessels, and a simulated tumor that resemble the human body. The moving part provides three independent movements, and by moving the pulmonary blood vessels, the simulated tumor, and the body surface unit, it can be used for simulation, analysis, and evaluation of dynamic tracking radiation therapy.

Features

  • The respiratory movements of the lung tumor, pulmonary blood vessels, and body surface unit can be operated independently.
  • Can be operated with a general-purpose PC or tablet (operated by Wi-Fi), so no communication cables are required.
  • Respiratory movements can be input with fixed waveforms (sine, cos4, cos6, triangular waves) and arbitrary waveforms (AZ-733VI).
    • Sub-millimeter accuracy of ±25mm.
    • Programs can be set with different cycles, amplitudes, and waveforms.
    • Programs can input respiratory movements using CSV files.

Waveform Pattern Example

Standard Waveform(cos4)

Arbitrary Input Waveform(AZ-733VI)

Specs

Body Dimensions680mm(W)×380mm(D)×200mm(H)
Pseudo-tumor, Pulmonary Vascular Movement±25mm
Body Surface Unit±10mm
Position Accuracy±0.2mm
Waveform Patternsin, cos4, cos6, Triangle wave, arbitrary waveform input(AZ-733VI)
Power Supply VoltageAC100V~AC240V 120VA
CommunicationWIFI